Recall Z-2260-2026 ยท OJK Eye Tray
Medline Industries, LP
Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P
| Date | May 28, 2026 (initiated Apr 10, 2026) |
| Status | Open, Classified |
| Reason | Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to it consignees on 4/10/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and |
| Quantity | 8 kits |
| Distribution | US Nationwide distribution in the states of California. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.