ClearanceBrief
Recall Z-2270-2026 ยท OFT Spinal Epidural Anesthesia Kit

B Braun Medical Inc

Brand Name: B. BRAUN MEDICAL INC. Product Name: SESK UPMC DOUBLE CATHETER TRAY Model/Catalog Number: 530134 Software Version: N/A Product Description: SESK UPMC DOUBLE CATHETER TRAY Component: No

DateMay 29, 2026 (initiated Jan 29, 2026)
StatusOpen, Classified
ReasonB. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
Root causeProcess control
ActionOn January 29, 2026 and May 6, 2026 URGENT MEDICAL DEVICE RECALL letters were sent to customers to notify them of a recall of a kit component. Actions to be taken include to review inventory to identify affected product, discontinue use and segregate. Using the acknowledgement form, respond to B Braun who will send further instructions to return the product. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or recalls@bbraunusa.com;
Quantity390
DistributionUS Nationwide distribution in the states of AL, FL, PA, TX, VA, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.