ClearanceBrief
Recall Z-2364-2026 ยท CAZ Anesthesia Conduction Kit

ARROW INTERNATIONAL, LLC

Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05060-PK; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/ASK-19609-VCU; CONTINUOUS NERVE BLOCK KIT/ASK-19600-CC2; CONTINUOUS NERV

DateJun 13, 2026 (initiated Apr 21, 2026)
StatusOpen, Classified
ReasonLidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Root causeNonconforming Material/Component
ActionOn 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device No
Quantity205,685
DistributionUS Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K00158719 GA PLASTIC REINFORCED EPIDURAL CATHETER
Arrow Intl., Inc., Jul 12, 2000
K103658FLEXTIP PLUS CLOSED TIP EPIDURAL CATHERTER
Teleflex, Inc., May 16, 2012
K121403CONTINUOUS PERIPHERAL NERVE BLOCK CATHETER KIT AND SET
Teleflex, Inc., Jun 20, 2012
K122027ARROW FLEXBLOCK CONTINUOUS PERIPHERAL NERVE BLOCK CATHETER
Arrow International, Inc., Aug 28, 2012
K801912ARROW POSITIVE PLACEMENT CON/EPIDURAL AN
Arrow Intl., Inc., Sep 16, 1980
K884552CONTINUOUS EPIDURAL ANESTHESIA KIT W/POLY CATHETER
Arrow Intl., Inc., Mar 14, 1989