Recall Z-2262-2026 ยท NEP System/Device, Pharmacy Compounding

Baxter Healthcare Corporation

Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

DateMay 29, 2026 (initiated Apr 22, 2026)
StatusOpen, Classified
ReasonAffected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
Root causeUnder Investigation by firm
ActionOn April 22, 2026, the firm began notifying consignees via letters titled URGENT PRODUCT RECALL. Customers were instructed to locate impacted product. Customers may choose to inspect impacted inlets and continue to use those with the correct 3-micron filter, or they may choose to discard impacted inlet lots without inspection and product will be replaced.
Quantity105675
DistributionWorldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Caroli
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002705EXACTA-MIX 2400 COMPOUNDING SYSTEM, MODEL EXACTA-MIX 2400
Baxa Corp., Mar 28, 2001