Recall Z-2259-2026 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029),

DateMay 28, 2026 (initiated Apr 30, 2026)
StatusOpen, Classified
ReasonIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Root causeUnder Investigation by firm
ActionOn April 30, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case motorized movements are unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2) and rotational (3) stand movements remain available. 3. In case the affected system has been transferred to another organization, please sen
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czec
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102005ALLURA XPER OR TABLE SERIES
Philips Medical Systems North America, Inc., Aug 9, 2010
K130638ALLURACLARITY XPER FD SERIES X-RAY SYSTEM
Philips Medical Systems Nederland B.V., Jun 28, 2013
K130842ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
Philips Medical Systems Nederland B.V., May 21, 2013
K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014
K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014
K161563Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Jul 29, 2016