Recall Z-2225-2026 · OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 722229. 2. Azurion 3 M15; Model Numbers: 722222, 722064, 722280, 722230. 3. Azurion 5 M12; Model Numbers: 722227, 722231. 4. Azurion 5 M20; Model Numbers: 722228, 722232, 722281. 5. Azurion 7 B12; M

DateMay 21, 2026 (initiated Apr 28, 2026)
StatusOpen, Classified
ReasonPotential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Root causeDevice Design
ActionPhilips Medical notified consignees on about 04/28/2026 via certified mailed letter titled "URGENT Medical Device Correction." Consignees were made aware of the issue and instructed to circulate the notification to all users of the system for awareness, that affected systems may continue to be used in accordance with their Instructions for Use (IFU), if the required position cannot be achieved using the Float Tabletop control for table movement, use alternative system movements (e.g., stand movement) where clinically feasible to adjust the relative position between the system and the patient,
Quantity5,537 units
DistributionNationwide distribution. International distribution to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, C¿te D'Ivoire, El Salvador, Estonia, Greece, Honduras, Hong Kong, Iceland, Indonesia, Ira
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020