ClearanceBrief
Recall Z-1963-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America

Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808

DateApr 27, 2026 (initiated Apr 14, 2026)
StatusOpen, Classified
ReasonThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Root causeUnder Investigation by firm
ActionPhilips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the pr
Quantity103 units
DistributionWorldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei
Philips Medical Systems Nederland B.V., Mar 3, 2022
K251808Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ing
Philips Medical Systems Nederland B.V., Jul 11, 2025