ClearanceBrief
Recall Z-1961-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America

Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.

DateApr 27, 2026 (initiated Apr 14, 2026)
StatusOpen, Classified
ReasonThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Root causeUnder Investigation by firm
ActionPhilips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the pr
Quantity86 units
DistributionWorldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Apr 10, 2020
K213583Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., May 16, 2022
K223458Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
Philips Medical Systems Nederland B.V., Apr 6, 2023
K251397Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR 5300 and MR 7700 MR System
Philips Medical Systems Nederland B.V., Jun 4, 2025
K251808Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ing
Philips Medical Systems Nederland B.V., Jul 11, 2025