Recall Z-2449-2026 ยท FSY Light, Surgical, Ceiling Mounted
Steris Corporation
HarmonyAIR A-Series Surgical Lighting System
| Date | Jun 15, 2026 (initiated May 15, 2026) |
| Status | Open, Classified |
| Reason | The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach. |
| Root cause | Material/Component Contamination |
| Action | An URGENT MEDICAL DEVICE RECALL notification letter dated 5/1/26 was sent to customers. STERIS Action All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their lighting system(s) to perform the correction as soon as possible. For the paint issue, each affected lighting system will undergo a full visual inspection for any signs of paint delamination; as needed a service kit will be installed to prevent paint chipping from occurring or if deemed necessary replaced. For the button issue, each affected lighting system will undergo inspectio |
| Quantity | 2101 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |