ClearanceBrief
Recall Z-2158-2026 ยท MSD Catheter, Hemodialysis, Implanted

ARROW INTERNATIONAL, LLC

Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM

DateMay 13, 2026 (initiated Apr 10, 2026)
StatusOpen, Classified
Reason16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
Root causeNonconforming Material/Component
ActionOn 4/10/2026, recall notices were mailed to customers who were informed of the following Arrow International LLC is recalling product due to the Merit Medical Systems, Inc. recall of the Merit 16F Dual-Valved Splitable Sheath Introducer component, which is included in impacted Arrow Hemodialysis Kits & Sets. Customers were asked to do the following: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Complete and return the Acknowledgement Form, see link: https://fca.myteleflex.com/en/recall/000605 3) This notice should be passed
Quantity34,568
DistributionWorldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111117CANNON II PLUS CHRONIC HEMODIALYSIS CATHETER, EDGE CHRONIC HEMODIALYSIS CATHETER, NEXTSTEP(R) RETROGRADE CHRONIC...
Teleflexmedical, Inc., Jun 30, 2011