Recall Z-2203-2026 ยท DYG Catheter, Flow Directed

Becton, Dickinson and Company

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P

DateMay 15, 2026 (initiated Apr 8, 2026)
StatusOpen, Classified
ReasonCatheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Root causeProcess change control
ActionOn 4/8/2026, mailed correction notices were mailed to customers and distributors who were informed of the following: Consider the following to reduce the potential for hub breakage issues. Minimize manipulation of proximal injectate hub and lumen. Pay frequent attention to the (blue) proximal injectate lumen for leakage or breakage, and exchange with another catheter if the issue is encountered. Modify existing clinical workflows to minimize flexing and stress of (blue) proximal injectate lumen. For example: o Securing the catheter lumens to an IV arm board to mitigate bending/flexing of
Quantity183
DistributionWorldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231248Swan-Ganz Catheters, FloTrac sensors, ClearSight finger cuffs, HemoSphere Advanced Monitoring Platform
Edwards Lifesciencess, LLC, Sep 21, 2023