Recall Z-2150-2026 ยท JAK System, X-Ray, Tomography, Computed

GE Medical Systems, LLC

Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6

DateMay 8, 2026 (initiated Mar 26, 2026)
StatusOpen, Classified
ReasonGE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Root causeSoftware design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not
Quantity38 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133705REVOLUTION CT
Ge Medical Systems, LLC, Apr 11, 2014
K163213Revolution CT
Ge Medical Systems, LLC, Dec 16, 2016
K191777Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Ge Medical Systems, LLC, Jul 26, 2019
K213715Revolution Apex, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select,Revolution CT; Revolution CT ES; Re
Ge Medical Systems, LLC, Dec 17, 2021