Recall Z-2149-2026 ยท JAK System, X-Ray, Tomography, Computed

GE Medical Systems, LLC

Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300

DateMay 8, 2026 (initiated Mar 26, 2026)
StatusOpen, Classified
ReasonGE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Root causeSoftware design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not
Quantity25 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213938Revolution Ascend
Ge Healthcare Japan Corporation, Feb 4, 2022