ClearanceBrief
Recall Z-2712-2026 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Elekta, Inc.

Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial therapy.

DateJul 9, 2026 (initiated Apr 3, 2026)
StatusOpen, Classified
ReasonDue to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
Root causeProcess control
ActionOn 04/03/2026, an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via email informing customers of a safety concern regarding detachment of the tip of an Intrauterine tube (IU tube) with a 4 mm diameter. Customers are instructed to: To prevent any potential health risks related to a detached tip, we emphasize the importance of adhering to the Instructions for Use: " Do not use an IU tube beyond its 3-year life expectancy or 300 reprocessing cycles, whichever comes first. " During the treatment preparation phase, before insertion of the applicator: Visua
Quantity1317
DistributionWorldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201272Geneva
Nucletron B.V., Jul 16, 2020