Recall Z-2274-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.

DateMay 29, 2026 (initiated Feb 4, 2026)
StatusOpen, Classified
ReasonPotential for the Cassette Loading Lever to break.
Root causeDevice Design
ActionFresenius Kabi notified consignees on about 02/04/2026 via letter. Consignees were informed of the lever breakage and that two potential sources of the breakage were: 1. If the Cassette Loading Lever is left open during transport, it can be damaged if it is jarred by another object or surface. 2. Applying force to a locked Loading Lever when it is in the locked position can result in potential damage and/or cause it to break. Consignees were reminded of proper Loading Lever handling: 1. Do not force the Cassette Loading Lever open or closed. The lever does not require significant force to
Quantity18,444 installed bases
DistributionUS Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MA, MD, ME, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019