Recall Z-2166-2026 ยท HIG Insufflator, Hysteroscopic

Stryker Corporation

Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are intended to be used with the FluidSmart Fluid Management System, which is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.

DateMay 12, 2026 (initiated Apr 9, 2026)
StatusOpen, Classified
ReasonDue to nonconforming products being inadvertently distributed.
Root causeProcess change control
ActionOn 04/09/2026, the firm sent via USPS Certified Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers that Stryker identified that certain lots of FluidSmart tube sets were manufactured with an incorrect pressure relief valve specification. This defect may prevent the pressure relief valve from activating at the intended pressure in the event of a capital device over pressure failure. One lot (Lot B000576601, quantity 10) impacted by this nonconformance was inadvertently shipped to a customer. Customers are instructed to: 1. Inform individuals within their organization who need to be aw
Quantity10 tube sets
DistributionU.S. Nationwide distribution in the state of GA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172048FluidSmart
Thermedx, LLC, Sep 5, 2017