ClearanceBrief
Recall Z-2042-2026 ยท LRP Tray, Surgical

Medline Industries, LP

Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670

DateApr 30, 2026 (initiated Mar 16, 2026)
StatusOpen, Classified
ReasonMedline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Root causeUnder Investigation by firm
ActionMedline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/16/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the affected product be quarantined pending additional labeling. Consignees were directed to respond to the notice via Website link: https://recalls.medline.com, and upon receipt of the submitted response form, the account will receive over-labels to place on affected inventory, with instructions for the user to remove and discard the affected component prior to using the kit. These labels will b
Quantity1590 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.