Recall Z-2041-2026 ยท LRO General Surgery Tray

Medline Industries, LP

Medline and Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. FOLEY CARE KIT, Medline Kit SKU UROT1044; 2. IV START KIT WITH EXTENSION TUBING, Medline Kit SKU IV8770; 3. IV START KIT, Medline Kit SKU DYNDV2520; 4. ARTERIAL LINE BUNDLE, Medline Kit SKU ART255; 5. ON/OFF TRAY, Medline Kit SKU DYNDC2969; 6. IV START KIT, Medline Kit SKU DYNDV2520; 7

DateApr 30, 2026 (initiated Mar 16, 2026)
StatusOpen, Classified
ReasonMedline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Root causeUnder Investigation by firm
ActionMedline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/16/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the affected product be quarantined pending additional labeling. Consignees were directed to respond to the notice via Website link: https://recalls.medline.com, and upon receipt of the submitted response form, the account will receive over-labels to place on affected inventory, with instructions for the user to remove and discard the affected component prior to using the kit. These labels will b
Quantity40845 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213481Medline General Surgery Tray
Medline Industries, Inc., May 24, 2022