ClearanceBrief
Recall Z-1971-2026 ยท OES Cardiac Catheterization Kit

AVID Medical, Inc.

Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.

DateMay 1, 2026 (initiated Mar 13, 2026)
StatusOpen, Classified
ReasonPotential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root causeUnder Investigation by firm
ActionOwens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The locat
Quantity180 kits
DistributionUS distribution to FL, IL, MO, MS, NC, NE, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.