Recall Z-1653-2026 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

DateApr 7, 2026 (initiated Mar 10, 2026)
StatusOpen, Classified
ReasonDuring patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Root causeRadiation Control for Health and Safety Act
ActionSiemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.
QuantityU.S. 758, OUS 2077
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241572ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono (VE40A) (ceiling); ARTIS pheno (VE40A)
Siemens Medical Solutions USA, Inc., Oct 22, 2024