Recall Z-2063-2026 ยท OCX Endoscopic Irrigation/Suction System

Erbe Medical, LLC

ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

DateMay 1, 2026 (initiated Mar 31, 2026)
StatusOpen, Classified
ReasonDevices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Root causeComponent change control
ActionOn 3/31/2026, correction notices were emailed to Medical Directors, Risk Managers, Medical Device Safety Officers, or Healthcare Professionals informing them to take the following actions: 1) Discontinue use of unsupported third-party Scope Port Connectors. 2) Provide this notice to all personnel within your organization who use, service, purchase, stock, or otherwise manage. 3) If you further distributed any affected devices, promptly forward this notice to those accounts and request that they take the same actions. 4) Continue following the instructions for use, including verifying connecto
Quantity23,359
DistributionUS Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221177Erbe's Tubing/Cap Sets
Erbe USA, Inc., Aug 11, 2022