Recall Z-1946-2026 ยท JDZ Mixer, Cement, For Clinical Use

Zimmer, Inc.

Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A

DateApr 24, 2026 (initiated Mar 30, 2026)
StatusOpen, Classified
ReasonTen complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Root causeVendor change control
ActionOn March 30, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Risk Managers were instructed to notify affected personnel and locate and quarantine affected product in inventory. A Zimmer Biomet sales representative may remove and return the affected product on your behalf, or the customer may choose to directly return all affected product from their facility. Surgeons were informed that there are no specific patient-monitoring instructions recommended beyond existing follow up schedules.
Quantity66960
DistributionUS Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.