Recall Z-2033-2026 ยท OJG Neurological Tray
Medline Industries, LP
Medline medical procedure kits, containing Medline Neuro Sponges (961 in total), NEURO,
| Date | Apr 30, 2026 (initiated Mar 19, 2026) |
| Status | Open, Classified |
| Reason | During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin. |
| Root cause | Under Investigation by firm |
| Action | Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their c |
| Quantity | 538201 kits |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.