ClearanceBrief
Recall Z-2019-2026 ยท OFX General Purpose Dental Tray

Medline Industries, LP

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920D; 2. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920F; 3. MEM MAXILLOFACIAL PACK-LF, Medline Kit Number/SKU DYNJ0660040S; 4. ORAL SURGERY PACK, Medline Kit Number/SKU DYNJ66450B; 5. ORAL SURGERY PACK RFID, Medline Kit Number/SKU DYNJ66450C; 6. ORAL

DateApr 30, 2026 (initiated Mar 19, 2026)
StatusOpen, Classified
ReasonDuring an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Root causeUnder Investigation by firm
ActionMedline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their c
Quantity2228 kits
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.