Recall Z-2008-2026 ยท HBA Neurosurgical Paddie
Medline Industries, LP
Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 0.5", XR, 10/PK, Medline SKU NEUROSPNG05; 3. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG06; 4. Neuro Sponges, Ep
| Date | Apr 30, 2026 (initiated Mar 19, 2026) |
| Status | Open, Classified |
| Reason | During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin. |
| Root cause | Under Investigation by firm |
| Action | Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their c |
| Quantity | 4225160 units |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K140369 | MEDLINE NEURO SPONGE Medline Industries, Inc., Sep 26, 2014 |