Recall Z-1918-2026 ยท IZL System, X-Ray, Mobile

Micro-X Ltd.

Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71

DateApr 22, 2026 (initiated Mar 27, 2026)
StatusOpen, Classified
ReasonX-Ray generator may malfunction resulting in the X-Ray being inoperable.
Root causeComponent design/selection
ActionOn March 27, 2026, Micro-X issued a Urgent: Medical Device Correction Notification, via E-Mail. Micro-X ask consignees to take the following actions: " Contact your local service representative to arrange replacement of the PCBA with the updated version at the next scheduled visit or Preventative Maintenance. In the interim, customers may continue to use the device. " To minimize the likelihood of system failure, the batteries should not be permitted to fully discharge. Any system errors or abnormal behaviours must be reported to your authorised service provider without delay. " If the produ
Quantity32 units
DistributionUS Nationwide distribution in the states and territories of WA, NJ, CA, Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211423Rover
Micro-X Limited, May 21, 2021