Recall Z-2097-2026 ยท OGR Ear, Nose, And Throat Surgical Tray
Medline Industries, LP
See RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD A
| Date | May 7, 2026 (initiated Jan 7, 2026) |
| Status | Open, Classified |
| Reason | Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. |
| Root cause | Process control |
| Action | An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 |
| Quantity | 11575 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.