ClearanceBrief
Recall Z-2094-2026 ยท OFF Central Venous Catheter Tray

Medline Industries, LP

Medline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268C; 4) TRAY 6 CENTRAL LINE A, Model Number:ACC010487; 5) PEDS PICC INSERTION TRAY, Model Number:CVI3650; 6) LINE ATTIRE KIT, Model Number:DYKM2170; 7) PK BNDL CENT LINE PROTECT KIT, Model Number:DYKM2274; 8)

DateMay 7, 2026 (initiated Jan 7, 2026)
StatusOpen, Classified
ReasonMedline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1
Quantity5562 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.