Recall Z-2091-2026 ยท OEQ Angiography/Angioplasty Kit

Medline Industries, LP

Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AA; 2) ENDO KIT W/SYRINGE, Model Number:DYKE1743; 3) BGMC ANGIOGRAPHY PACK-LF, Model Number:DYNJ0800934D; 4) GENERAL ENDO PACK-LF, Model Number:DYNJ0842873J; 5) ANGIO PROCEDURE PACK, Model Number:DYNJ32328; 6) OR ANGIO PACK-LF, Model Number:DYNJ43415C; 7) ANGIO PACK II, Model Number:DYNJ44066A; 8) SOUTH CATH LAB PACK,

DateMay 7, 2026 (initiated Jan 7, 2026)
StatusOpen, Classified
ReasonMedline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1
Quantity6961 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.