Recall Z-2165-2026 ยท QBJ Integrated Continuous Glucose Monitoring System, Factory Calibrated

Dexcom, Inc.

Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

DateMay 11, 2026 (initiated Apr 14, 2026)
StatusOpen, Classified
ReasonA software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic eve
Root causeSoftware design
ActionOn 04/14/2026, the firm sent via the iOS App an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" communication informing customers that a software error was identified within app versions 1.5 to 1.6 and G7 iOS / G7 Watch iOS app v2.9.0 through v2.11.2 that can cause the app display to flicker sequentially through previous estimated glucose values (EGVs) and trend arrows before displaying the current EGV and trend arrow with the potential to delay glucose alerts. Required Customer Actions: Updating from app versions 1.5-1.6 to the most current version is mandatory and you will not be able to u
Quantity70212
DistributionSW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC,
FDA recordOpen on fda.gov