Recall Z-2001-2026 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Abbott Point Of Care Inc.
i-STAT G3+ cartridge; List Number: 03P78-26;
| Date | Apr 29, 2026 (initiated Apr 1, 2026) |
| Status | Open, Classified |
| Reason | Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, po |
| Root cause | Process change control |
| Action | On April 1, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. This Urgent Medical Device Corrective is to customers in the U.S. with the following recommended actions: " Discontinue use of the affected i-STAT EG7+, EG6+, and G3+ cartridge lots for reporting PCO and pH results. " Use an alternate cartridge type, alternate lot, or testing method. " Complete and return the Business Reply Card included with this notification to confirm receipt and understanding. If you have forwarded any of the specified i-STAT EG7+, EG6+ or G3+ cartridges to another facility, we request that |
| Quantity | 40275 units (US only) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K223857 | i-STAT G3+ cartridge with the i-STAT 1 System Abbott Point of Care, Inc., Sep 15, 2023 |