Recall Z-2552-2026 ยท HCG Device, Neurovascular Embolization

Stryker Neurovascular

INZONE DETACHMENT SYSTEM, REF: M00345100950

DateJun 24, 2026 (initiated Apr 6, 2026)
StatusOpen, Classified
ReasonEmbolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
Root causeComponent design/selection
ActionOn 4/3/2026, recall notices were mailed to Supply Chain Management and Recall Coordinators who were asked to do the following: 1) Segregate the affected devices in a secure location for return to recalling firm. 2) If any of the subject devices have been distributed to other organizations, please forward a copy of this notice to the new users and provide contact details to recalling firm so new recipients can be informed appropriately. 3) Maintain awareness of this communication internally until all required actions have been completed within your facility. 4) Complete and return the Busines
Quantity44,937
DistributionWorldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212455InZone Detachment System
Stryker Neurovascular, Nov 16, 2021