Recall Z-1823-2026 · DQO Catheter, Intravascular, Diagnostic

Gentuity, LLC

Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;

DateApr 16, 2026 (initiated Mar 23, 2026)
StatusOpen, Classified
ReasonThe potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
Root causeDevice Design
ActionOn March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
Quantity184 units (28 US, 156 OUS)
DistributionDomestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242966Gentuity® HF-OCT Imaging System with Vis-Rx Prime® Micro-Imaging Catheter
Gentuity, LLC, Jan 31, 2025