Recall Z-1910-2026 ยท OWL Peripheral Catheter Insertion Kit

ARROW INTERNATIONAL, LLC

Arrow Echogenic Introducer Needle REF ASK-04001-BWK2 UDI code: (01)10801902220315(17)261231(11)251210(10)33F25J0347 This device is intended for injection or aspiration of fluids. The needle protection device covers the needle after use to help prevent needle sticks.

DateApr 22, 2026 (initiated Mar 11, 2026)
StatusOpen, Classified
ReasonDue to receiving a notice from a supplier that the liquid adhesive were incorrectly manufactured
Root causeProcess control
ActionOn 03/18/2026, the firm sent via FedEx-2-day an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that Ferndale Laboratories issued a Field Safety Notice (FSN) to inform users of multiple complaints reporting the butyrate tubing of the Mastisol Liquid Adhesive vial cracked during actuation. According to the Ferndale Laboratories FSN, one lot of MASTISOL Liquid Adhesive was exposed to a double sterilization cycle which may have contributed to the butyrate tubing becoming brittle and susceptible to cracking during actuation. Medical Facilities should: 1. Immediately check inventory for
Quantity85
DistributionU.S. Nationwide distribution in the states of AZ, CA, GA, MA, ME, MI, OR, PA, and VA.
FDA recordOpen on fda.gov