ClearanceBrief
Recall Z-1889-2026 ยท KKX Drape, Surgical, Antimicrobial

Medline Industries, LP

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes E P DRAPE PACK-LF DYNJ0373061J

DateApr 17, 2026 (initiated Feb 27, 2026)
StatusOpen, Classified
ReasonUnapproved design changes to the products outside of the 510(k) clearance.
Root causeComponent change control
ActionOn February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Quantity270,311 total
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020393MASTER & FRANK SURGICAL DRAPES (STERILE)
Master & Frank Enterprses Co., Ltd., May 3, 2002