ClearanceBrief
Recall Z-1876-2026 ยท OEQ Angiography/Angioplasty Kit

Medline Industries, LP

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes JUDKINS PACK DYNJ51126 ANGIO PACK DYNJ32555C, DYNJ66631 ANGIO TRAY DYNJ46153B ANGIOGRAPHY PACK DYNJ26855F ANGIOGRAPHY PACK-M-LF DYNJ24530I ANGIOGRAPHY TRAY-LF DYNJ0220880V, DYNJ30077D CATH LAB ANGIO PACK-LF DYNJ0545064X CATH LAB PACK DYNJ39057C FEMORAL ANGIOGRAPY SET UP DYNJ61015B RAD-ANGIO P

DateApr 17, 2026 (initiated Feb 27, 2026)
StatusOpen, Classified
ReasonUnapproved design changes to the products outside of the 510(k) clearance.
Root causeComponent change control
ActionOn February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Quantity270,311 total
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.