Recall Z-1722-2026 ยท FDE Laparoscopy Kit
Medline Industries, LP
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.
| Date | Apr 7, 2026 (initiated Feb 27, 2026) |
| Status | Open, Classified |
| Reason | Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient a |
| Quantity | 20 kits |
| Distribution | Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK). |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.