Recall Z-1716-2026 ยท OEQ Angiography/Angioplasty Kit

Medline Industries, LP

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 00-HCT055P; 2. INTERVENTIONAL CATH LAB PK-LF, Medline SKU DYNJ0376230Q; 3. ANGIOGRAPHY CATH LAB, Medline SKU DYNJ23532C; 4. ANGIO LEFT HEART PACK, Medline SKU DYNJ40632C; 5. ANGIO DRAPE PACK, Medline SKU DYNJ47952B;

DateApr 7, 2026 (initiated Feb 27, 2026)
StatusOpen, Classified
ReasonMedline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root causeUnder Investigation by firm
ActionMedline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient a
Quantity17902 kits
DistributionWorldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).