Recall Z-1713-2026 ยท DXT Injector And Syringe, Angiographic

Medline Industries, LP

NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,R

DateApr 7, 2026 (initiated Feb 27, 2026)
StatusOpen, Classified
ReasonMedline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root causeUnder Investigation by firm
ActionMedline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient a
Quantity192690 units
DistributionWorldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113198NMI CONTROL SYRINGE, 7 ML NMI CONTROL SYRINGE, 8 ML
Navilyst Medical, Inc., Nov 28, 2011