Recall Z-1825-2026 ยท JAK System, X-Ray, Tomography, Computed

PHILIPS MEDICAL SYSTEMS

Philips Spectral CT on Rails. Model Number: 728334.

DateApr 16, 2026 (initiated Mar 7, 2026)
StatusOpen, Classified
ReasonPhilips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? or should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Root causeUnder Investigation by firm
ActionPhilips Medical notified consignees on about 03/07/2026 via letter. Consignees were instructed to identify any affected systems in inventory and specifically for each issue: 1. Issue #1: Ensure to follow the instructions mentioned in the IFU Section 4 Preparing for an Examination and Sub section Gantry Operation. 2. Issue #2: Avoid the Select Previous Surview Unavailable issue, manually remove the patient's middle name while importing the patient detail from HIS/RIS. 3. Issue #3: if Define Head Area option is grayed out with DoseRight Index (DRI) on, review the dose levels displayed on the U
Quantity3 systems
DistributionWorldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212875Spectral CT on Rails
Philips Medical Systems Nederland B.V., May 13, 2022