ClearanceBrief
Recall Z-2046-2026 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to suppor

DateMay 1, 2026 (initiated Apr 2, 2026)
StatusOpen, Classified
ReasonDue to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Root causeNonconforming Material/Component
ActionOn 04/02/2026, the firm sent via 2-day mailing service an "Urgent Medical Device Correction" letter to customers informing them that Intuitive identified that the screws on the cannula mount of the USM from a specific population are susceptible to breaking over time. When the cannula mount experiences a breakage of all six screws, the mount can separate from the rest of the USM. Breakage of any fewer than six screws will not result in cannula mount separation, or impact to clinical performance. How to recognize issue/failure: Screw breakage or cannula mount loosening may be observable as visi
Quantity2781
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Au
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131861DA VINCI SURGICAL SYSTEM, ENDOWRIST INSTRUMENTS AND ACCESSORIES
Intuitive Surgical, Inc., Mar 28, 2014
K171294da Vinci X Surgical System
Intuitive Surgical, Inc., May 26, 2017