Recall Z-1942-2026 ยท NOU Continuous, Ventilator, Home Use

Ventec Life Systems, Inc.

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

DateMay 1, 2026 (initiated Mar 25, 2026)
StatusOpen, Classified
ReasonRespiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Root causeProcess change control
ActionOn 3/25/2026 recall notices were mailed to customers who were asked to do the following: 1. Immediately discontinue use of affected devices. 2. Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3. Use a non-affected ventilator for patient support. 4. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5. Complete and return the response form via email
Quantity42
DistributionUS: OH, GA, CA, MO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162877VOCSN Unified Respiratory System
Ventec Life Systems, Inc., Apr 7, 2017