Recall Z-1772-2026 ยท LTT Panels, Test, Susceptibility, Antimicrobial

Beckman Coulter, Inc.

MicroScan Neg MIC 3J REF C54814

DateApr 3, 2026 (initiated Jun 25, 2025)
StatusOpen, Classified
ReasonDue to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Root causeProcess control
ActionOn 07/15/2024, firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customer inform them that customers have reported multiple patient samples with skipped wells for Ceftolozane/Tazobactam (C/T) when tested with MicroScan Neg MIC 3J Lot 2025-08-23. The C/T 2/4 dilution well exhibited no growth while growth was observed in the C/T 4/4 well. Customers are instructed to: "Discontinue use of this lot and discard per your laboratory guidelines. "Consider reviewing the C/T antimicrobial susceptibility test results from this lot if Enterobacterales reported as resistant and P. aerugi
Quantity688 boxes (20 panels per box)
DistributionInternational distribution in the country of Japan.
FDA recordOpen on fda.gov