ClearanceBrief
Recall Z-1802-2026 ยท FJK Set, Tubing, Blood, With And Without Anti-Regurgitation Valve

B Braun Medical Inc

B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.

DateApr 15, 2026 (initiated Mar 19, 2026)
StatusOpen, Classified
ReasonPotential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Root causeDevice Design
ActionB. Braun notified consignees on about 03/19/2026 via letter titled URGENT MEDICAL DEVICE CORRECTION. Consignees were instructed to use alternative dialysis equipment and disposables, if possible. If not, BBMI recommends the following actions that can help mitigate potential harm should you need to use sets impacted by this field action on patients: 1. Ensure full adherence to both the bloodline and dialysis machine Instructions for Use (IFU). a. Streamline bloodlines - strict adherence to the priming process as outlined in the IFU, with special emphasis on Step 26. b. B3 bloodlines - strict
Quantity4,848 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080807BLOOD TUBING SETS (STERILE FLUID PATH)
Nxstage Medical, Inc., Jul 18, 2008