ClearanceBrief
Recall Z-1851-2026 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Olympus Corporation of the Americas

Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm. Model Number: TB2-0520FC.

DateApr 17, 2026 (initiated Mar 23, 2026)
StatusOpen, Classified
ReasonPotential for detachment of a distal tip component of the device during use.
Root causeDevice Design
ActionOlympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were transferred or further distributed.
Quantity830 units
DistributionInternational distribution in the countries of Australia, Japan, Hong Kong, and Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K252150Ultrasonic Bipolar Generator (USG-410); Reuseable Cordless Transducer (TD-410); THUNDERBEAT II Shears with Ultrasonic Mo
Olympus Medical Systems Corporation, Oct 7, 2025