Recall Z-1616-2026 ยท CBK Ventilator, Continuous, Facility Use

Philips Respironics, Inc.

Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.

DateMar 26, 2026 (initiated Mar 2, 2026)
StatusOpen, Classified
ReasonUsing non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Root causeUnder Investigation by firm
ActionAn UPDATED: "URGENT FIELD SAFETY NOTICE" dated 3/2/26 was mailed to consignees containing recall information for three different failure modes that can be experienced in all Trilogy Evo, Trilogy EVO O2, Trilogy Evo Universal, and Trilogy EV300 devices. The firm identified that using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. This can lead to under-delivery of therapy to the patient. Consignees are being instructed to update all devices with software version 1.05.15.00 thro
Quantity113,717 units
DistributionWorldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.