Recall Z-1691-2026 ยท PIO Enteral Specific Transition Connectors

Medline Industries, LP

Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT

DateMar 27, 2026 (initiated Feb 13, 2026)
StatusOpen, Classified
ReasonConnectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Root causeProcess change control
ActionOn February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.
Quantity9,040 kits
DistributionWorldwide - US Nationwide and the countries of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.