TMJ Bilateral Implants, REF: CHG020

DateJun 22, 2026 (initiated Feb 11, 2026)
StatusOpen, Classified
ReasonCustom temporomandibular joint implant may have contained incorrect patient-specific components that led to a poor fit of the implant, leading to a right-sided open bite.
Root causeProcess control
ActionOn 2/11/2026, the customer was contacted and verbally informed of the recall. On 4/23/2026, a recall notice was sent to the customer asking them to take the following action: 1) Inform all appropriate members of your organization who need to be made aware of this action. 2) The affected product is already implanted. Following explantation, during revision surgery, the explanted device shall be returned to the firm's sales representative. 3) Complete and return the Business Reply Form (BRF) via email to cmf-pfa@stryker.com Please contact the firm's quality department cmf-pfa@stryker.com wit
Quantity1
DistributionUS Nationwide distribution in the state of MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.