Recall Z-1893-2026 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

(1)(2) LeadCare II Blood Lead Test Kit, (3)LeadCare II 5X Blood Lead Test Kit, (4)LeadCare II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 pa

DateApr 17, 2026 (initiated Mar 13, 2026)
StatusOpen, Classified
ReasonCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Root causeDevice Design
ActionOn March 13, 2026, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. REQUIRED ACTIONS: This field action does NOT require removal of product from the field. The LeadCare Test Systems continue to perform as intended when used according to the product labeling and Instructions for Use. The regulatory status of all products noted above remains unchanged and the product supply uninterrupted. Technical and customer support will continue as normal. LEADCARE II: The only blood collection devices validated for use with the Magellan LeadCare II Blood Lead Test Kit are the capillary tubes
Quantity197893
DistributionWorldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, Sou
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052549LEADCARE II BLOOD LEAD TESTING SYSTEM
Esa Biosciences, Inc., Oct 6, 2005