Recall Z-1576-2026 ยท MSD Catheter, Hemodialysis, Implanted
Merit Medical Systems, Inc.
ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
| Date | Mar 26, 2026 (initiated Feb 13, 2026) |
| Status | Open, Classified |
| Reason | 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access. |
| Root cause | Process control |
| Action | On 2/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: A picture of the 16F Dual-Valved Splittable Sheath Introducer is attached to help you identify the product. Place this notice and the picture on or near the affected product in your inventory (if applicable). Discontinue the use of the 16F Dual-Valved Splittable Sheath Introducer and destroy the sheath introducer at the point of use. Ensure that applicable personnel within your organization are made aware of this field action. If the product has been further distributed to other facilities |
| Quantity | 131 |
| Distribution | Worldwide distribution: US (nationwide) to states of: TX, LA, OR, NM, MI, FL, CO, CA, GA, NC, AK, NY, PA, WI, NJ, VA, AZ, IN, KY, MO, DE, SC, MT, NE, WV, OH, MA, MN, CT, MD, KS, WA, TN, ID, UT, DC, MS, IL, OK, AL, RI, AR, NV and OUS (International) to countries of : Canada, Spain, United Arab Emirat |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K090148 | PROGUIDE CHRONIC DIALYSIS CATHETER Merit Medical Systems, Inc., Apr 13, 2009 |